1. పరిచయం
The Extech 382252 is a comprehensive Earth Ground Resistance Tester Kit designed for accurate measurement of earth ground resistance. This kit provides three selectable ranges for earth ground resistance: 20 Ohms (Ω), 200 Ohms (Ω), and 2000 Ohms (Ω). Beyond earth ground testing, the device also functions as a multimeter, capable of measuring AC/DC Voltage, Resistance, and Continuity. It features automatic zero adjustment for ease of use. This manual provides detailed instructions for the safe and effective operation, maintenance, and troubleshooting of your Extech 382252.
2. భద్రతా సమాచారం
హెచ్చరిక: Improper use of this meter can cause damage, shock, injury, or death. Read and understand all safety information and operating instructions before using this instrument. Always adhere to local and national safety codes.
- ఏదైనా ఫంక్షన్ కోసం గరిష్ట ఇన్పుట్ పరిమితులను మించకూడదు.
- Exercise extreme caution when working with live electrical circuits.
- Ensure the meter is in good working condition and free from damage before use.
- కొలతల కోసం ఎల్లప్పుడూ సరైన ఫంక్షన్ మరియు పరిధిని ఉపయోగించండి.
- ఖచ్చితమైన రీడింగ్లను నిర్ధారించడానికి తక్కువ బ్యాటరీ సూచిక కనిపించినప్పుడు బ్యాటరీలను మార్చండి.
- Keep hands and fingers away from test probe tips when making measurements.
- పేలుడు వాతావరణాలలో లేదా మండే వాయువులు లేదా ధూళి సమక్షంలో మీటర్ను ఆపరేట్ చేయవద్దు.
3. ప్యాకేజీ విషయాలు
అన్ప్యాక్ చేసిన తర్వాత అన్ని వస్తువులు ఉన్నాయని మరియు పాడవకుండా ఉన్నాయని ధృవీకరించండి:
- Extech 382252 Earth Ground Resistance Tester
- Test Leads (set)
- Auxiliary Earth Bars (2 pieces)
- క్యారీయింగ్ కేసు
- Protective Rubber Holster
- 1.5V AA బ్యాటరీలు (6 ముక్కలు)

Image 1: The Extech 382252 Earth Ground Resistance Tester Kit, showing the main unit, test leads, two auxiliary earth bars, and the carrying case.
4. ఉత్పత్తి లక్షణాలు
The Extech 382252 offers the following key features:
- Earth Ground Resistance measurement in three ranges: 20Ω, 200Ω, and 2000Ω.
- AC/DC వాల్యూమ్tage measurement capabilities.
- Resistance and Continuity testing functions.
- Automatic zero adjustment for accurate readings.
- పెద్ద, సులభంగా చదవగలిగే LCD డిస్ప్లే.
- Data Hold function to freeze displayed readings.
- తక్కువ కాంతి పరిస్థితుల్లో దృశ్యమానత కోసం బ్యాక్లైట్.
- Durable protective rubber holster.

చిత్రం 2: ముందు view of the Extech 382252 Earth Ground Resistance Tester, highlighting its display, function dial, and control buttons.
5. సెటప్
5.1 బ్యాటరీ ఇన్స్టాలేషన్
- మీటర్ పవర్ ఆఫ్ చేయబడిందని నిర్ధారించుకోండి.
- మీటర్ వెనుక భాగంలో బ్యాటరీ కంపార్ట్మెంట్ కవర్ను గుర్తించండి.
- Use a screwdriver to remove the screw(s) securing the cover.
- Insert six (6) 1.5V AA batteries, observing correct polarity (+/-).
- బ్యాటరీ కంపార్ట్మెంట్ కవర్ను మార్చి, దానిని స్క్రూ(లు)తో భద్రపరచండి.
5.2 టెస్ట్ లీడ్లను కనెక్ట్ చేయడం
Connect the test leads to the appropriate input jacks on the meter for the desired measurement function. For earth ground resistance testing, specific connections involving the auxiliary earth bars will be detailed in the operating instructions.
6. ఆపరేటింగ్ సూచనలు
6.1 పవర్ ఆన్/ఆఫ్
Rotate the function dial from the 'OFF' position to any desired measurement function to power on the meter. To power off, rotate the dial back to the 'OFF' position.
6.2 Automatic Zero Adjustment
The meter features automatic zero adjustment. When selecting a resistance range, the meter will automatically adjust to zero, compensating for lead resistance. Ensure test leads are not shorted or open during this process for accurate zeroing.
6.3 డేటా హోల్డ్ ఫంక్షన్
Press the 'HOLD' button to freeze the current reading on the display. Press it again to release the hold and resume live readings.
6.4 బ్యాక్లైట్ ఫంక్షన్
Press the backlight button (often indicated by a light bulb icon) to illuminate the display for better visibility in dark environments. Press it again to turn off the backlight.
7. Earth Ground Resistance Measurement
The Extech 382252 uses the 3-wire (or Fall-of-Potential) method for earth ground resistance measurements. This method requires the use of the two auxiliary earth bars provided.
7.1 Measurement Procedure (3-Wire Method)
- తయారీ: Ensure the circuit under test is de-energized and safely isolated.
- Placement of Auxiliary Earth Bars:
- Drive the first auxiliary earth bar (P) into the ground at a suitable distance from the earth electrode (E) being tested.
- Drive the second auxiliary earth bar (C) into the ground at a further distance from P, ensuring all three points (E, P, C) are in a straight line. Typical distances are 5-10 meters (15-30 feet) between E and P, and P and C, but may vary based on soil conditions and electrode size.
- టెస్ట్ లీడ్లను కనెక్ట్ చేస్తోంది:
- Connect the 'E' terminal of the meter to the earth electrode under test.
- Connect the 'P' terminal of the meter to the first auxiliary earth bar (P).
- Connect the 'C' terminal of the meter to the second auxiliary earth bar (C).
- పరిధిని ఎంచుకోండి: Rotate the function dial to the desired Earth Ground Resistance range (20Ω, 200Ω, or 2000Ω). Start with the highest range if the resistance value is unknown.
- Perform Test: Press the 'TEST' button. The meter will inject a current and measure the voltage drop to calculate the earth ground resistance.
- కొలత చదవండి: The resistance value will be displayed on the LCD.
- Verify Accuracy: For critical measurements, it is recommended to repeat the test with the auxiliary earth bars placed at slightly different distances to ensure the reading is stable and accurate.
8. Other Measurements
8.1 AC/DC వాల్యూమ్tagఇ కొలత
- Connect the red test lead to the 'V' input jack and the black test lead to the 'COM' input jack.
- Rotate the function dial to the desired ACV (750V AC) or DCV (1000V DC) range.
- కొలవవలసిన సర్క్యూట్ లేదా భాగం అంతటా పరీక్ష ప్రోబ్లను కనెక్ట్ చేయండి.
- సంపుటాన్ని చదవండిtagడిస్ప్లేలో ఇ విలువ.
8.2 నిరోధక కొలత
- సర్క్యూట్ లేదా భాగం డి-ఎనర్జైజ్ చేయబడిందని నిర్ధారించుకోండి.
- Connect the red test lead to the 'V' input jack and the black test lead to the 'COM' input jack.
- Rotate the function dial to the desired Resistance range (e.g., 200Ω, 20kΩ, 200kΩ).
- పరీక్ష ప్రోబ్లను కాంపోనెంట్ అంతటా కనెక్ట్ చేయండి.
- ప్రదర్శనలో నిరోధక విలువను చదవండి.
8.3 కొనసాగింపు పరీక్ష
- సర్క్యూట్ లేదా భాగం డి-ఎనర్జైజ్ చేయబడిందని నిర్ధారించుకోండి.
- Connect the red test lead to the 'V' input jack and the black test lead to the 'COM' input jack.
- Rotate the function dial to the Continuity position (often indicated by a speaker icon).
- Connect the test probes across the component or wire.
- An audible beep indicates continuity (low resistance). The display will show the resistance value.
9. నిర్వహణ
9.1 శుభ్రపరచడం
Wipe the meter with a dry, soft cloth. Do not use abrasives or solvents. Periodically clean the test leads and probes to ensure good electrical contact.
9.2 బ్యాటరీ భర్తీ
డిస్ప్లేలో తక్కువ బ్యాటరీ సూచిక కనిపించినప్పుడు, కొలత ఖచ్చితత్వాన్ని నిర్వహించడానికి బ్యాటరీలను వెంటనే మార్చండి. బ్యాటరీ ఇన్స్టాలేషన్ సూచనల కోసం విభాగం 5.1 చూడండి.
9.3 నిల్వ
If the meter is to be stored for an extended period, remove the batteries to prevent leakage and damage. Store the meter in its carrying case in a cool, dry environment, away from direct sunlight and extreme temperatures.
10. ట్రబుల్షూటింగ్
- No Display/Meter Not Powering On: Check battery installation and ensure batteries are fresh. Verify the function dial is not in the 'OFF' position.
- సరికాని రీడింగ్లు: Ensure test leads are properly connected and not damaged. Check battery level. For earth ground resistance, verify proper placement and connection of auxiliary earth bars and ensure good contact with the ground.
- నిరంతర బీప్ లేదు: Check if the circuit is truly continuous. Ensure test leads are making good contact.
- Display Shows 'OL' (Overload): The measured value exceeds the selected range. Switch to a higher range.
If issues persist after following these steps, contact customer support.
11. స్పెసిఫికేషన్లు
| స్పెసిఫికేషన్ | విలువ |
|---|---|
| బ్రాండ్ | Extech |
| మోడల్ సంఖ్య | 382252 |
| Earth Ground Resistance Ranges | 20Ω, 200Ω, 2000Ω |
| AC వాల్యూమ్tagఇ పరిధి | 750V వరకు AC |
| DC సంtagఇ పరిధి | 1000V DC వరకు |
| ప్రతిఘటన శ్రేణులు | 20Ω, 200Ω, 20kΩ, 200kΩ (typical, check device for full ranges) |
| కొలత ఖచ్చితత్వం | +/-0.5% (for certain functions, refer to product manual for specifics) |
| శక్తి మూలం | 6 x 1.5V AA బ్యాటరీలు |
| వస్తువు బరువు | 0.69 కిలోలు (1.52 పౌండ్లు) |
| ఉత్పత్తి కొలతలు (L x W x H) | 7.9 x 3.6 x 2 అంగుళాలు |
| చేర్చబడిన భాగాలు | Test leads, 2 earth bars, carrying case, holster, six 1.5V AA batteries |
| భద్రతా రేటింగ్ | CAT III 1000V, CE |
12. వారంటీ
The Extech 382252 Earth Ground Resistance Tester Kit typically comes with a 1-సంవత్సరం పరిమిత వారంటీ from the date of purchase. This warranty covers defects in materials and workmanship under normal use. It does not cover damage resulting from misuse, accident, unauthorized repair, or modification. Please retain your proof of purchase for warranty claims.
13. మద్దతు
For technical assistance, calibration services, or further information regarding your Extech 382252, please refer to the official Extech website or contact their customer support department. Contact details can usually be found on the manufacturer's webసైట్ లేదా మీ ఉత్పత్తితో చేర్చబడిన డాక్యుమెంటేషన్లో.