The Extech FL700 Fluoride Meter is a precision instrument designed for the rapid and accurate measurement of fluoride ions in various aqueous samples, including drinking water. This device integrates a sensing electrode, measurement electronics, and a digital display into a single, portable unit, making it suitable for both field and laboratory applications.

The FL700 provides direct readings of total fluoride concentrations ranging from 9.99 ppm or mg/L down to 0.1 ppm. It features automatic temperature compensation to ensure measurement accuracy across varying environmental conditions.

2. భద్రతా సమాచారం

Please read and understand all safety precautions before operating the Extech FL700 Fluoride Meter. Failure to follow these instructions may result in injury or damage to the instrument.

  • Always handle the fluoride electrode with care. Avoid touching the sensing element directly.
  • When working with TISAB reagent tablets or any chemical solutions, wear appropriate personal protective equipment, including gloves and eye protection.
  • Do not immerse the entire meter in liquid. The meter is designed for electrode immersion only.
  • Store the meter and reagents in a cool, dry place, away from direct sunlight and extreme temperatures.
  • పిల్లలకు దూరంగా ఉంచండి.

3. ఉత్పత్తి ముగిసిందిview

The Extech FL700 Fluoride Meter is a compact and user-friendly device. Key components and features are outlined below:

Extech FL700 Fluoride Meter front view
ముందు view of the Extech FL700 Fluoride Meter, showing the display and control buttons.
Hand holding Extech FL700 Fluoride Meter
The Extech FL700 Fluoride Meter held in hand, demonstrating its ergonomic design and clear display.
Close-up of Extech FL700 electrode tip
క్లోజ్-అప్ view of the fluoride electrode tip, which is integrated into the meter.
Extech FL700 Fluoride Meter in use with sample
The Extech FL700 Fluoride Meter being used to measure a liquid sample, showing the meter in its stand.
Extech FL700 Fluoride Meter and included accessories
All components included with the Extech FL700: the meter, sensor cap, TISAB reagent tablets, and neckstrap.

ముఖ్య లక్షణాలు:

  • రెండు పాయింట్ల క్రమాంకనం: Ensures accurate measurements across the fluoride range.
  • మెమరీ నిల్వ: Stores up to 25 labeled readings for later review.
  • Field or lab testing: Designed for versatile use in various environments.
  • Fast response: Provides fluoride measurements in less than 1 minute.
  • Large dual LCD: Simultaneously displays ppm (or mg/L) and temperature with a bargraph.

4. సెటప్

4.1. బ్యాటరీ ఇన్‌స్టాలేషన్

The Extech FL700 requires four 3V button batteries (included). To install or replace batteries:

  1. మీటర్ పైభాగంలో ఉన్న బ్యాటరీ కంపార్ట్‌మెంట్ క్యాప్‌ను తిప్పండి.
  2. Insert the four 3V button batteries, ensuring correct polarity as indicated inside the compartment.
  3. Replace the battery compartment cap and twist to secure.

4.2. ప్రారంభ తయారీ

Before first use or after prolonged storage, prepare the electrode:

  1. Remove the sensor cap from the electrode tip.
  2. ఎలక్ట్రోడ్ కొనను డిస్టిల్డ్ వాటర్ తో శుభ్రం చేసుకోండి.
  3. Condition the electrode by immersing it in a fluoride standard solution (e.g., 1 ppm or 10 ppm) for approximately 15-30 minutes.

5. ఆపరేటింగ్ సూచనలు

5.1. పవర్ చేయడం ఆన్/ఆఫ్

నొక్కండి ఆన్/ఆఫ్ button to turn the meter on or off.

5.2. Calibration (Two-Point)

Accurate calibration is crucial for reliable measurements. The FL700 supports two-point calibration.

  1. Prepare two fluoride standard solutions, typically one low concentration (e.g., 1 ppm) and one high concentration (e.g., 10 ppm). Ensure the solutions are at a stable temperature.
  2. స్వేదనజలంతో ఎలక్ట్రోడ్‌ను కడిగి ఆరబెట్టండి.
  3. Immerse the electrode into the first standard solution.
  4. నొక్కండి మరియు పట్టుకోండి CAL / RECALL button until 'CAL' appears on the display. The meter will automatically recognize the standard and calibrate. Wait for the reading to stabilize.
  5. Rinse the electrode again and immerse it into the second standard solution.
  6. నొక్కండి మరియు పట్టుకోండి CAL / RECALL button again. The meter will complete the two-point calibration.
  7. The meter is now calibrated and ready for measurement.

5.3. కొలత తీసుకోవడం

To measure fluoride concentration in a sampలే:

  1. మీటర్ క్రమాంకనం చేయబడిందని నిర్ధారించుకోండి.
  2. స్వేదనజలంతో ఎలక్ట్రోడ్‌ను కడిగి ఆరబెట్టండి.
  3. Add TISAB (Total Ionic Strength Adjustment Buffer) reagent to your sample according to the reagent instructions. TISAB ensures that the fluoride ions are in a measurable form and minimizes interference.
  4. Immerse the electrode tip into the prepared sample solution. Ensure the tip is fully submerged.
  5. Allow the reading to stabilize on the display. The large dual LCD will show the fluoride concentration in ppm (or mg/L) and the temperature.
  6. ప్రస్తుత రీడింగ్‌ను పట్టుకోవడానికి, మోడ్ / హోల్డ్ బటన్. విడుదల చేయడానికి మళ్ళీ నొక్కండి.

5.4. Recalling Stored Readings

The meter can store up to 25 labeled readings.

  • To recall stored readings, press the CAL / RECALL button briefly when not in calibration mode.
  • ఉపయోగించండి మోడ్ / హోల్డ్ button to scroll through the stored readings.

6. నిర్వహణ

6.1. ఎలక్ట్రోడ్ క్లీనింగ్

Regular cleaning of the electrode ensures optimal performance.

  • After each use, rinse the electrode thoroughly with distilled water.
  • If the electrode becomes fouled or readings become erratic, clean it with a mild detergent solution or a specialized electrode cleaning solution. Rinse thoroughly with distilled water afterward.

6.2. నిల్వ

Proper storage extends the life of the electrode and meter.

  • Always replace the sensor cap on the electrode tip when not in use. The cap should contain a small amount of electrode storage solution or a fluoride standard solution to keep the membrane hydrated. Do not store dry.
  • మీటర్‌ను చల్లని, పొడి ప్రదేశంలో నిల్వ చేయండి.

6.3. బ్యాటరీ భర్తీ

Replace batteries when the low battery indicator appears on the display or when the meter fails to power on. Refer to Section 4.1 for battery installation instructions.

7. ట్రబుల్షూటింగ్

సమస్యసాధ్యమైన కారణంపరిష్కారం
మీటర్ పవర్ ఆన్ చేయడం లేదుడెడ్ లేదా తప్పుగా ఇన్‌స్టాల్ చేయబడిన బ్యాటరీలుబ్యాటరీ ధ్రువణతను తనిఖీ చేయండి; బ్యాటరీలను భర్తీ చేయండి.
అనియత లేదా అస్థిర రీడింగ్‌లుDirty electrode; air bubbles on electrode; insufficient sample volume; expired or incorrect TISAB.Clean electrode; gently tap electrode to remove bubbles; ensure electrode is fully submerged; check TISAB.
Meter fails to calibrateIncorrect standard solutions; dirty electrode; electrode not conditioned; damaged electrode.Verify standard concentrations; clean and condition electrode; replace electrode if damaged.
తక్కువ బ్యాటరీ సూచికబ్యాటరీలు తక్కువగా ఉన్నాయి.Replace all four 3V button batteries.

8. స్పెసిఫికేషన్లు

స్పెసిఫికేషన్వివరాలు
తయారీదారుExtech
పార్ట్ నంబర్FL700
వస్తువు బరువు7.7 ఔన్సులు
ఉత్పత్తి కొలతలు6.8 x 1.6 x 1.4 అంగుళాలు
అంశం మోడల్ సంఖ్యFL700
బ్యాటరీలు4 లిథియం మెటల్ బ్యాటరీలు అవసరం (చేర్చబడినవి)
శైలిFL700
అంశం ప్యాకేజీ పరిమాణం1
ముక్కల సంఖ్య1
చేర్చబడిన భాగాలుUnit, Instruction Guide
బ్యాటరీ సెల్ రకంఆల్కలీన్
మొదట అందుబాటులో ఉన్న తేదీజూలై 15, 2008

9. వారంటీ సమాచారం

The Extech FL700 Fluoride Meter comes with a full one-year warranty on parts and workmanship, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modification.

10. మద్దతు

For technical support, service, or additional information, please visit the official Extech webసైట్‌లో లేదా వారి కస్టమర్ సర్వీస్ విభాగాన్ని సంప్రదించండి. మీరు మరింత సమాచారం మరియు ఉత్పత్తులను కూడా ఇక్కడ కనుగొనవచ్చు అమెజాన్‌లో ఎక్స్‌టెక్ స్టోర్.

సంబంధిత పత్రాలు - FL700

ముందుగాview EXTECH 45170 4-in-1 తేమ, ఉష్ణోగ్రత, గాలి ప్రవాహం మరియు కాంతి మీటర్ వినియోగదారు మాన్యువల్
EXTECH 45170 కోసం యూజర్ మాన్యువల్, ఇది తేమ, ఉష్ణోగ్రత, గాలి ప్రవాహం మరియు కాంతిని కొలిచే 4-ఇన్-1 మీటర్. ఆపరేషన్, స్పెసిఫికేషన్లు మరియు వారంటీ సమాచారాన్ని కలిగి ఉంటుంది.
ముందుగాview రేడియాజియోన్ UV-AB ఎక్స్‌టెక్ UV505 నుండి మాన్యువల్ యుటెంటే మిసురాటోర్
రేడియోజియోన్ UV-AB ఎక్స్‌టెక్ UV505 ద్వారా మిసురాటోర్ ద్వారా మాన్యువల్ పూర్తి అవుతుంది. ఇస్ట్రుజియోని ఆపరేటివ్, స్పెసిఫిక్ టెక్నిచ్, ఇన్ఫర్మేజియోని డి సిక్యూరెజా మరియు మాన్యుటెన్జియోన్‌ను చేర్చండి.
ముందుగాview Extech UV505 UV-AB లైట్ మీటర్ యూజర్ మాన్యువల్
ఎక్స్‌టెక్ UV505 UV-AB లైట్ మీటర్ కోసం యూజర్ మాన్యువల్. ఈ పత్రం UVA మరియు UVB లైట్‌ను కొలవడానికి పరికరం యొక్క లక్షణాలు, ఆపరేషన్, స్పెసిఫికేషన్‌లు, భద్రతా జాగ్రత్తలు మరియు నిర్వహణపై వివరణాత్మక సమాచారాన్ని అందిస్తుంది.
ముందుగాview ఎక్స్‌టెక్ EX810 1000 Amp Klemmeter IR థర్మామీటర్ Gebruikershandleiding కలుసుకున్నారు
Gedetailleerde gebruikersandleiding voor de Extech EX810 1000 Amp klemmeter మెట్ ingebouwde IR థర్మామీటర్. బెవాట్ ఇన్‌స్ట్రక్షన్‌లు, వీలిఘీడ్‌స్రిచ్ట్‌లిజ్‌నెన్ ఎన్ టెక్నీస్ స్పెసిఫికేటీస్ వోర్ నావ్‌కేరిజ్ మెటింగెన్ వ్యాన్ స్పానింగ్, స్ట్రూమ్, వీర్‌స్టాండ్, కెపాసిటీట్, ఫ్రీక్వెన్సీ మరియు టెంపరేచర్.
ముందుగాview ఎక్స్‌టెక్ EX650 సిరీస్ యూజర్ గైడ్: 600A ట్రూ RMS డిజిటల్ Clamp మీటర్లు
600A ట్రూ RMS డిజిటల్ క్లాక్ యొక్క Extech EX650 సిరీస్ కోసం సమగ్ర వినియోగదారు గైడ్amp మీటర్లు, EX650 మరియు EX655 మోడళ్లను కవర్ చేస్తాయి. ఖచ్చితమైన AC/DC వాల్యూమ్ కోసం ఆపరేషన్, ఫీచర్లు, భద్రత మరియు స్పెసిఫికేషన్ల గురించి తెలుసుకోండి.tage, కరెంట్, రెసిస్టెన్స్ మరియు మరిన్ని కొలతలు.
ముందుగాview ఎక్స్‌టెక్ DM100 పాకెట్ మల్టీమీటర్ యూజర్ గైడ్
ఎక్స్‌టెక్ DM100 4000 కౌంట్ ఆటోరేంజింగ్ పాకెట్ మల్టీమీటర్ కోసం యూజర్ గైడ్, భద్రత, స్పెసిఫికేషన్లు, ఆపరేషన్, ఫీచర్లు మరియు నిర్వహణను కవర్ చేస్తుంది.