Extech SDL150 కరిగిన ఆక్సిజన్ మీటర్ SD లాగర్ యూజర్ మాన్యువల్

మోడల్: SDL150

1. పరిచయం

The Extech SDL150 is a Dissolved Oxygen Meter with an integrated SD card data logger. This device is designed for precise measurement of dissolved oxygen (DO) concentration and temperature in various applications. It records data directly onto an SD card in Excel format for easy transfer and analysis on a personal computer.

Key features include a dual backlit display, automatic temperature compensation, offset adjustment for relative measurements, and user-programmable sampలింగ్ రేట్లు.

Extech SDL150 Dissolved Oxygen Meter with probe

మూర్తి 1: Extech SDL150 Dissolved Oxygen Meter and Probe. This image shows the main unit with its digital display and control buttons, connected to the dissolved oxygen probe.

2. ప్యాకేజీ కంటెంట్‌లు మరియు సెటప్

2.1. ప్యాకేజీ విషయాలు

Verify that all items are present and in good condition upon unpacking. The standard package includes:

  • Extech SDL150 Dissolved Oxygen Meter
  • Dissolved Oxygen Probe with built-in temperature sensor
  • 6 x AA బ్యాటరీలు
  • SD card (2G capacity)
  • Two spare replacement membranes for the DO probe
  • Electrolyte solution for the DO probe
  • హార్డ్ క్యారీయింగ్ కేసు

2.2. ప్రారంభ సెటప్

  1. బ్యాటరీ ఇన్‌స్టాలేషన్: Open the battery compartment on the back of the meter and insert the 6 AA batteries, observing correct polarity.
  2. SD కార్డ్ చొప్పించడం: Locate the SD card slot, typically on the side of the meter. Insert the provided 2G SD card until it clicks into place.
  3. Probe Preparation:
    • Unscrew the protective cap from the dissolved oxygen probe.
    • Fill the probe's membrane cap with the supplied electrolyte solution. Ensure no air bubbles are trapped.
    • Carefully screw the membrane cap back onto the probe, ensuring the membrane is intact and free of damage.
  4. ప్రోబ్ కనెక్షన్: Connect the prepared DO probe to the designated input port on the top of the SDL150 meter.
  5. పవర్ ఆన్: నొక్కండి శక్తి పరికరాన్ని ఆన్ చేయడానికి బటన్.

3. ఆపరేటింగ్ సూచనలు

3.1. ప్రాథమిక కొలతలు

  1. క్రమాంకనం: Before first use and periodically thereafter, calibrate the dissolved oxygen probe according to the instructions in the full user manual (not provided here). This typically involves air calibration or zero-oxygen solution calibration.
  2. ఇమ్మర్షన్: Immerse the DO probe into the sample solution. Ensure the membrane is fully submerged and free from air bubbles.
  3. రీడింగ్ డిస్‌ప్లే: The dual backlit display will show the dissolved oxygen concentration (in mg/L or %) and the temperature (in °F or °C).
  4. Units Selection: ఉపయోగించండి ఫంక్షన్ button to cycle through different measurement units (mg/L, %, °F, °C) if applicable.
Extech SDL150 Dissolved Oxygen Meter in use in a laboratory setting

మూర్తి 2: Extech SDL150 in a laboratory environment. The meter is shown measuring dissolved oxygen in a beaker, demonstrating its application.

3.2. డేటా లాగింగ్

  • Manual Logging: నొక్కండి లాగ్ button to manually store the current reading. The meter can store up to 99 manual readings.
  • Automatic Logging:
    • Access the settings menu (refer to the full manual for specific button combinations) to set the desired sampలింగ్ రేటు.
    • ఒకసారి ఎస్ampling rate is set, activate automatic logging. The meter will record up to 20 million readings on the 2G SD card.
    • Data is stored in Excel-compatible format for easy transfer to a PC via the SD card.

3.3. అదనపు విధులు

  • డేటా హోల్డ్: నొక్కండి పట్టుకోండి button to freeze the current display reading. Press again to release.
  • కనిష్ట/గరిష్టం: ఉపయోగించండి MIN / MAX బటన్ view the minimum and maximum recorded values during a measurement session.
  • బ్యాక్‌లైట్: The display features a backlight for improved visibility in low-light conditions. Activate it using the dedicated backlight button (often indicated by a light bulb icon).
  • ఆటో పవర్ ఆఫ్: The meter includes an auto power-off function to conserve battery life. This feature can typically be disabled or adjusted in the settings.
  • ఆఫ్‌సెట్ సర్దుబాటు: The offset adjustment function allows for zeroing the meter for relative measurements. Consult the full manual for detailed instructions on using this feature.

4. నిర్వహణ

4.1. Probe Membrane and Electrolyte Replacement

The dissolved oxygen probe requires periodic maintenance, specifically the replacement of the membrane and electrolyte solution. This ensures accurate readings and extends the life of the probe.

  1. Unscrew the membrane cap from the probe body.
  2. Carefully remove the old membrane and discard it.
  3. Rinse the probe tip and the inside of the new membrane cap with distilled water.
  4. Fill the new membrane cap with fresh electrolyte solution, ensuring no air bubbles are present.
  5. Gently screw the new membrane cap onto the probe body. Avoid overtightening.
  6. Allow the probe to stabilize for a few minutes before use and perform a calibration.

4.2. శుభ్రపరచడం మరియు నిల్వ చేయడం

  • మీటర్ క్లీనింగ్: మీటర్ యొక్క బాహ్య భాగాన్ని మృదువైన, d తో తుడవండి.amp గుడ్డ. రాపిడి క్లీనర్లు లేదా ద్రావకాలు ఉపయోగించవద్దు.
  • ప్రోబ్ క్లీనింగ్: After each use, rinse the probe tip with distilled water. Store the probe with its protective cap on, ideally with a small amount of electrolyte solution to keep the membrane moist, as specified in the full manual.
  • దీర్ఘ-కాల నిల్వ: For extended storage, remove the batteries from the meter and ensure the probe is properly prepared as per the full manual's instructions. Store the unit and accessories in the hard carrying case in a cool, dry place.

5. ట్రబుల్షూటింగ్

This section provides general guidance for common issues. For detailed troubleshooting steps, refer to the comprehensive user manual provided with your Extech SDL150.

  • శక్తి లేదు: బ్యాటరీ ఇన్‌స్టాలేషన్‌ను తనిఖీ చేయండి మరియు బ్యాటరీలు ఖాళీ కాలేదని నిర్ధారించుకోండి. అవసరమైతే భర్తీ చేయండి.
  • సరికాని రీడింగ్‌లు:
    • క్రమాంకనం చేయండి.
    • Check the probe membrane for damage or air bubbles. Replace the membrane and electrolyte if needed.
    • Ensure the probe is properly connected to the meter.
  • SD కార్డ్ రికార్డ్ కావడం లేదు:
    • Verify the SD card is correctly inserted.
    • Check if the SD card is full. Transfer data to a PC and clear the card if necessary.
    • Ensure the logging function is activated and the sampling rate is set.

6. స్పెసిఫికేషన్లు

పరామితివిలువ
కరిగిన ఆక్సిజన్ పరిధి0 to 20.0 mg/L (0 to 100.0%)
ఉష్ణోగ్రత పరిధి0 నుండి 50°C (32 నుండి 122°F)
కొలత ఖచ్చితత్వం+/-0.5%
ఉష్ణోగ్రత పరిహారంAutomatic, 0 to 50°C
డేటా లాగింగ్ సామర్థ్యం99 manual readings, 20M readings via 2G SD card
డేటా ఫార్మాట్Excel compatible (.xls)
పిసి ఇంటర్ఫేస్Built-in (via SD card)
శక్తి మూలం6 x AA బ్యాటరీలు (చేర్చబడ్డాయి)
కొలతలు (L x W x H)7.2 x 2.9 x 1.9 అంగుళాలు (18.3 x 7.4 x 4.8 సెం.మీ.)
బరువు1.61 పౌండ్లు (0.73 కిలోలు)

7. వారంటీ మరియు మద్దతు

For information regarding warranty coverage, technical support, or service, please contact Extech customer service. Refer to the official Extech website or the documentation included with your product for current contact details and warranty terms.

తయారీదారు: Extech

సంబంధిత పత్రాలు - SDL150

ముందుగాview ఎక్స్‌టెక్ సమగ్ర కేటలాగ్: పర్యావరణ మీటర్లు, ఎనిమోమీటర్లు మరియు పరీక్షా పరికరాలు
ఎక్స్‌టెక్ యొక్క విస్తృత శ్రేణి అధిక-నాణ్యత పర్యావరణ మీటర్లు, థర్మో-ఎనిమోమీటర్లు, గ్యాస్ డిటెక్టర్లు, నీటి నాణ్యత పరీక్షకులు, సౌండ్ మీటర్లు, లైట్ మీటర్లు మరియు పారిశ్రామిక, HVAC/R మరియు ప్రొఫెషనల్ అప్లికేషన్‌ల కోసం రూపొందించబడిన ఇతర ఖచ్చితత్వ పరీక్ష మరియు కొలత పరికరాలను అన్వేషించండి.
ముందుగాview EXTECH 45170 4-in-1 తేమ, ఉష్ణోగ్రత, గాలి ప్రవాహం మరియు కాంతి మీటర్ వినియోగదారు మాన్యువల్
EXTECH 45170 కోసం యూజర్ మాన్యువల్, ఇది తేమ, ఉష్ణోగ్రత, గాలి ప్రవాహం మరియు కాంతిని కొలిచే 4-ఇన్-1 మీటర్. ఆపరేషన్, స్పెసిఫికేషన్లు మరియు వారంటీ సమాచారాన్ని కలిగి ఉంటుంది.
ముందుగాview రేడియాజియోన్ UV-AB ఎక్స్‌టెక్ UV505 నుండి మాన్యువల్ యుటెంటే మిసురాటోర్
రేడియోజియోన్ UV-AB ఎక్స్‌టెక్ UV505 ద్వారా మిసురాటోర్ ద్వారా మాన్యువల్ పూర్తి అవుతుంది. ఇస్ట్రుజియోని ఆపరేటివ్, స్పెసిఫిక్ టెక్నిచ్, ఇన్ఫర్మేజియోని డి సిక్యూరెజా మరియు మాన్యుటెన్జియోన్‌ను చేర్చండి.
ముందుగాview Extech UV505 UV-AB లైట్ మీటర్ యూజర్ మాన్యువల్
ఎక్స్‌టెక్ UV505 UV-AB లైట్ మీటర్ కోసం యూజర్ మాన్యువల్. ఈ పత్రం UVA మరియు UVB లైట్‌ను కొలవడానికి పరికరం యొక్క లక్షణాలు, ఆపరేషన్, స్పెసిఫికేషన్‌లు, భద్రతా జాగ్రత్తలు మరియు నిర్వహణపై వివరణాత్మక సమాచారాన్ని అందిస్తుంది.
ముందుగాview ఎక్స్‌టెక్ EX810 1000 Amp Klemmeter IR థర్మామీటర్ Gebruikershandleiding కలుసుకున్నారు
Gedetailleerde gebruikersandleiding voor de Extech EX810 1000 Amp klemmeter మెట్ ingebouwde IR థర్మామీటర్. బెవాట్ ఇన్‌స్ట్రక్షన్‌లు, వీలిఘీడ్‌స్రిచ్ట్‌లిజ్‌నెన్ ఎన్ టెక్నీస్ స్పెసిఫికేటీస్ వోర్ నావ్‌కేరిజ్ మెటింగెన్ వ్యాన్ స్పానింగ్, స్ట్రూమ్, వీర్‌స్టాండ్, కెపాసిటీట్, ఫ్రీక్వెన్సీ మరియు టెంపరేచర్.
ముందుగాview ఎక్స్‌టెక్ EX650 సిరీస్ యూజర్ గైడ్: 600A ట్రూ RMS డిజిటల్ Clamp మీటర్లు
600A ట్రూ RMS డిజిటల్ క్లాక్ యొక్క Extech EX650 సిరీస్ కోసం సమగ్ర వినియోగదారు గైడ్amp మీటర్లు, EX650 మరియు EX655 మోడళ్లను కవర్ చేస్తాయి. ఖచ్చితమైన AC/DC వాల్యూమ్ కోసం ఆపరేషన్, ఫీచర్లు, భద్రత మరియు స్పెసిఫికేషన్ల గురించి తెలుసుకోండి.tage, కరెంట్, రెసిస్టెన్స్ మరియు మరిన్ని కొలతలు.