1. పరిచయం
This manual provides essential instructions for the safe and effective operation of the Hioki 3030-10 HiTester Analog Multimeter. This device is an average-sensing, manual-ranging meter designed for measuring voltage, resistance, and current in various electrical testing applications. Please read this manual thoroughly before use and retain it for future reference.
2. భద్రతా సమాచారం
To ensure safe operation and prevent damage to the device, observe the following safety precautions:
- The meter is designed to meet International Electrotechnical Commission (IEC) safety standard 61010.
- It is certified for Category III installations up to 600V, suitable for use with enclosed building circuit installations (e.g., service panel parts, branch circuits, fixed installations connected directly to AC power supply).
- A 0.5-amp/250V fuse is installed to protect the operator from electric shock due to transient voltage. Do not bypass or use an incorrect fuse.
- Always select the appropriate measurement range before connecting the test leads to the circuit.
- ఏ ఫంక్షన్కైనా గరిష్ట ఇన్పుట్ రేటింగ్లను మించకూడదు.
- Inspect test leads for damage before each use. Replace if insulation is compromised.
- The meter is designed to withstand falls of up to 1 meter onto hardwood or concrete. However, avoid dropping the device.
3. ప్యాకేజీ విషయాలు
అన్ని అంశాలు ఉన్నాయని ధృవీకరించండి:
- Hioki 3030-10 HiTester Analog Multimeter
- Detachable Test Leads (Red and Black)
- క్యారీయింగ్ కేసు
- వినియోగదారు మాన్యువల్ (ఈ పత్రం)
4. ఉత్పత్తి ముగిసిందిview మరియు ఫీచర్లు

మూర్తి 1: Hioki 3030-10 HiTester Analog Multimeter. The image displays the multimeter unit, its protective carrying case, and the red and black test leads. The multimeter features a clear analog display with multiple scales for various measurements, a central rotary switch for function and range selection, and input jacks for the test leads.
The Hioki 3030-10 is an analog multimeter designed for reliable electrical measurements. Key features include:
- అనలాగ్ డిస్ప్లే: Provides a clear, real-time indication of measurements, useful for observing trends and fluctuations.
- Manual Ranging: Requires the user to select the appropriate measurement range for precision.
- Average-Sensing: Delivers accurate readings for linear loads with undistorted waveforms.
- కొలత విధులు: DC/AC వాల్యూమ్లను కొలుస్తుందిtagఇ, DC కరెంట్ మరియు రెసిస్టెన్స్.
- Battery Test Feature: Determines charge level for 1.5V batteries.
- LED Test Function: Allows testing of LED lights.
- Temperature Measurement Capability: Requires a separately sold thermistor probe.
- భద్రతా సమ్మతి: Meets IEC 61010, CAT III 600V safety standards.
- మన్నికైన డిజైన్: Engineered to withstand drops up to 1 meter.
5. సెటప్
5.1 బ్యాటరీ ఇన్స్టాలేషన్
The Hioki 3030-10 requires two AAA batteries for operation. To install or replace batteries:
- Ensure the multimeter is turned OFF (rotary switch set to 'OFF').
- యూనిట్ వెనుక భాగంలో బ్యాటరీ కంపార్ట్మెంట్ కవర్ను గుర్తించండి.
- Open the cover and insert two AAA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
- బ్యాటరీ కంపార్ట్మెంట్ కవర్ను సురక్షితంగా మూసివేయండి.
5.2 టెస్ట్ లీడ్ కనెక్షన్
Connect the detachable test leads to the multimeter:
- "COM" (సాధారణ) ఇన్పుట్ జాక్లోకి బ్లాక్ టెస్ట్ లీడ్ను చొప్పించండి.
- Insert the red test lead into the "VΩmA" input jack for voltage, resistance, and current measurements.
- Ensure connections are firm before proceeding with any measurements.
6. ఆపరేటింగ్ సూచనలు
The Hioki 3030-10 is a manual-ranging multimeter. Always select the appropriate function and range using the rotary switch before making a measurement.
6.1 DC వాల్యూమ్tagఇ కొలత (V)
- రోటరీ స్విచ్ను కావలసిన DC వాల్యూమ్కు సెట్ చేయండిtage (V) range (e.g., 0.3V, 3V, 12V, 30V, 120V, 300V, 600V). Start with the highest range if the voltagఇ తెలియదు.
- ఎరుపు పరీక్ష తాడును సర్క్యూట్ యొక్క సానుకూల (+) వైపుకు మరియు నలుపు పరీక్ష తాడును ప్రతికూల (-) వైపుకు కనెక్ట్ చేయండి.
- సంపుటాన్ని చదవండిtage value from the appropriate DCV scale on the analog display.
6.2 AC వాల్యూమ్tage Measurement (~V)
- రోటరీ స్విచ్ను కావలసిన AC వాల్యూమ్కు సెట్ చేయండిtage (~V) range (e.g., 12V, 30V, 120V, 300V, 600V). Start with the highest range if the voltagఇ తెలియదు.
- AC వాల్యూమ్ అంతటా టెస్ట్ లీడ్లను కనెక్ట్ చేయండి.tagఇ మూలం.
- సంపుటాన్ని చదవండిtage value from the appropriate ACV scale on the analog display.
6.3 DC Current Measurement (A)
జాగ్రత్త: కరెంట్ కొలిచేటప్పుడు ఎల్లప్పుడూ మల్టీమీటర్ను సర్క్యూట్తో సిరీస్లో కనెక్ట్ చేయండి. వాల్యూమ్ అంతటా సమాంతరంగా ఎప్పుడూ కనెక్ట్ చేయవద్దు.tagఇ మూలం.
- Set the rotary switch to the desired DC Current (A) range (e.g., 60µA, 30mA, 300mA). Start with the highest range if the current is unknown.
- Break the circuit and connect the multimeter in series, ensuring the current flows through the meter.
- Read the current value from the appropriate DCA scale on the analog display.
6.4 రెసిస్టెన్స్ మెజర్మెంట్ (Ω)
జాగ్రత్త: నిరోధకతను కొలిచే ముందు సర్క్యూట్ డీ-ఎనర్జైజ్ చేయబడిందని నిర్ధారించుకోండి.
- Set the rotary switch to the desired Resistance (Ω) range (e.g., x1, x10, x100, x1k).
- Before measuring, short the test leads together and adjust the "Ω ADJ" knob (if present) to zero the meter.
- కొలవవలసిన భాగం అంతటా పరీక్ష లీడ్లను కనెక్ట్ చేయండి.
- Read the resistance value from the Ω scale on the analog display and multiply by the selected range multiplier.
6.5 Battery Test (BATT 1.5V)
- Set the rotary switch to the "BATT 1.5V" position.
- Connect the red test lead to the positive (+) terminal of the 1.5V battery and the black test lead to the negative (-) terminal.
- Read the battery charge level from the "BATT" scale on the analog display.
6.6 LED Test (LED)
- Set the rotary switch to the "LED" position.
- Connect the test leads to the LED. Observe polarity; if the LED does not light, reverse the leads.
- A lit LED indicates functionality.
6.7 ఉష్ణోగ్రత కొలత (TEMP)
This function requires a separately purchased thermistor probe. Refer to the thermistor probe's manual for specific connection and operation instructions.
- Set the rotary switch to the "TEMP" position.
- Connect the thermistor probe to the designated input jacks.
- Read the temperature from the "TEMP(°C)" scale on the analog display.
7. నిర్వహణ
7.1 శుభ్రపరచడం
మీటర్ను తుడవండి casing తో a సాఫ్ట్, damp cloth. Do not use abrasive cleaners or solvents. Ensure the device is off and disconnected from all circuits before cleaning.
7.2 ఫ్యూజ్ రీప్లేస్మెంట్
If the current measurement function ceases to operate, the fuse may need replacement. The Hioki 3030-10 uses a 0.5-amp/250V fuse. Always replace with a fuse of the identical type and rating.
- మల్టీమీటర్ ఆఫ్ చేయబడిందని మరియు పరీక్ష లీడ్లు డిస్కనెక్ట్ చేయబడిందని నిర్ధారించుకోండి.
- Refer to the device's internal diagram (if available) or contact customer support for instructions on accessing the fuse compartment.
- Carefully remove the old fuse and insert the new 0.5A/250V fuse.
- Reassemble the meter securely.
7.3 నిల్వ
When not in use, store the multimeter in its carrying case in a cool, dry place, away from direct sunlight and extreme temperatures. Remove batteries if storing for extended periods to prevent leakage.
8. ట్రబుల్షూటింగ్
If the multimeter does not function as expected, consider the following:
- డిస్ప్లే/పవర్ లేదు: బ్యాటరీ ఇన్స్టాలేషన్ను తనిఖీ చేయండి మరియు బ్యాటరీలు ఖాళీ కాలేదని నిర్ధారించుకోండి. అవసరమైతే భర్తీ చేయండి.
- సరికాని రీడింగ్లు:
- Ensure the correct function and range are selected for the measurement.
- పరీక్ష లీడ్లు సరిగ్గా కనెక్ట్ అయ్యాయని మరియు దెబ్బతినలేదని ధృవీకరించండి.
- For resistance measurements, ensure the circuit is de-energized and the meter is zeroed.
- Current Function Not Working: Check the internal fuse (refer to Section 7.2).
- సూది కదలకపోవడం: Ensure the meter is powered on and the test leads are making good contact with the circuit. Check for open circuits.
If problems persist after attempting these steps, contact Hioki customer support for further assistance.
9. స్పెసిఫికేషన్లు
| పరామితి | విలువ |
|---|---|
| మోడల్ సంఖ్య | 3030-10 |
| కొలత రకం | Analog Multimeter (Average-Sensing, Manual-Ranging) |
| DC సంtagఇ పరిధులు | 0.3V, 3V, 12V, 30V, 120V, 300V, 600V |
| AC వాల్యూమ్tagఇ పరిధులు | 12V, 30V, 120V, 300V, 600V |
| DC కరెంట్ పరిధులు | 60µA, 30mA, 300mA |
| ప్రతిఘటన శ్రేణులు | x1, x10, x100, x1k (Max 3MΩ) |
| బ్యాటరీ పరీక్ష | 1.5V బ్యాటరీలు |
| ఫ్యూజ్ ప్రొటెక్షన్ | 0.5 ఎ / 250 వి |
| భద్రతా ప్రమాణం | IEC 61010, CAT III 600V |
| శక్తి మూలం | 2 x AAA బ్యాటరీలు (చేర్చబడలేదు) |
| వస్తువు బరువు | 150 గ్రా |
| ఉత్పత్తి కొలతలు (L x W x H) | 15.75 x 15.49 x 5.84 సెం.మీ |
| ఆపరేటింగ్ వాల్యూమ్tage (Min.) | 0.3 వోల్ట్లు (DC) |
| ఎగువ ఉష్ణోగ్రత రేటింగ్ | 150 డిగ్రీల సెల్సియస్ |